- The monochromator with a holographic flashing grating is used. It has the advantage of high wavelength precision, good monochromaticity and low stray light.
- A sample compartment for 0.5cm-10cm cells.
- Direct concentration read-out and concentration factor setting function.
- Equipped with RS232 port.
Technical Specifications |
|
Wavelength Range |
340 – 1000 nm |
Transmittance Range |
0 – 199.9 |
Absorbance Range A |
0 – 1.999 |
Scope of concentration display |
0 – 1999 |
Wavelength Accuracy (nm) |
±3 |
Wavelength Repeatability (nm) |
1.5 |
Transmittance Accuracy |
±1% |
Absorbance Repeatability |
0.3% |
Bandwidth |
2nm |
Stray Light |
1.0% |
Noise |
0% 0.3 |
Noise |
100% 1.0 |
Overall Dimensions |
460 x 350 x 180 |
Weight |
8 Kgs |
Display |
LCD Screen with backlight |
Output Mode |
Standard RS232 serial port |